System Verification - Tune (Detector Optimization) Portion
Instrument Name : MSD5973N_US21934101
DC Polarity : Positive
Filament : 1
BasePeak should be 69 or 219 OK
Position of mass 69 69.00 OK
Position of mass 219 219.00 OK
Position of mass 502 502.00 OK
Position of isotope mass 70 70.00 OK
Position of isotope mass 220 219.99 OK
Position of isotope mass 503 503.02 OK
Ratio of mass 70 to mass 69(0.5 - 1.6%) 1.08 OK
Ratio of mass 220 to mass 219(3.2 - 5.4%) 4.23 OK
Ratio of mass 503 to mass 502(7.9 - 12.3%) 10.28 OK
Ratio of 219 to 69 should be > 40% and is 85.19 OK
Ratio of 502 to 69 should be > 2.4% and is 5.24 OK
Mass 69 Precursor (<= 3%) 0.11 OK
Mass 219 Precursor (<= 6%) 0.20 OK
Mass 502 Precursor (<= 12%) 0.42 OK
Testing for a leak in the system
Ratio of 18 to 69 (<20%) 4.86 OK
Ratio of 28 to 69 (<10%) 3.61 OK
Electron Multiplier Voltage 1718 OK
Tune portion of System Verification passed.
System Verification

 

 

 

System Verification - Tune (Detector Optimization) Portion
Instrument Name : MSD5973N_US21934101
DC Polarity : Positive
Filament : 2
BasePeak should be 69 or 219 OK
Position of mass 69 69.00 OK
Position of mass 219 218.96 OK
Position of mass 502 502.00 OK
Position of isotope mass 70 70.01 OK
Position of isotope mass 220 219.99 OK
Position of isotope mass 503 503.01 OK
Ratio of mass 70 to mass 69(0.5 - 1.6%) 1.11 OK
Ratio of mass 220 to mass 219(3.2 - 5.4%) 4.27 OK
Ratio of mass 503 to mass 502(7.9 - 12.3%) 10.00 OK
Ratio of 219 to 69 should be > 40% and is 107.39 OK
Ratio of 502 to 69 should be > 2.4% and is 10.30 OK
Mass 69 Precursor (<= 3%) 0.20 OK
Mass 219 Precursor (<= 6%) 0.29 OK
Mass 502 Precursor (<= 12%) 0.50 OK
Testing for a leak in the system
Ratio of 18 to 69 (<20%) 5.17 OK
Ratio of 28 to 69 (<10%) 3.54 OK
Electron Multiplier Voltage 1671 OK
Tune portion of System Verification passed.
System